PIRANHA MULTI
Brand
:
RTI
Country:
Sweden
The Piranha MULTI X-ray meter is all you need for qualified diagnostics and QA on Rad/Fluoro, CT, Dental, and Mammography X-ray scans. Measures kVp, Time, HVL, Total Filtration, Dose, Dose rate, presents Waveform, and much more.
Customize your measurements and reports with Ocean Next™ software.
The unique feature of the RTI Piranha MULTI X-ray meter makes it possible to check the position of the detector before measuring. By using the Position Check function, you can verify that the detector area is fully irradiated. Possible field inhomogeneities are also neutralized.
Piranha MULTI — The complete solution for X-ray QA Testing!
Specifications for R/F
- Tube Voltage: 35 – 160 kV (±1.5 %)
- PPV: yes
- Dose: 1.3 nGy - 1500 Гр (±5%)
- Time: 0.1ms – 2000 s (±0.5 % or ±0.5 ms)
- Dose rate: 15 nGy/s – 320 mGy/s (±5 % or ±7 nGy/s)
- HVL: 0,72 – 13 mm Al (±10 % or ±0.2 mm)
- Total Filtration: 1.0 – 90 mm Al (±10 % or ±0.3 mm)
- Pulses: 1 – 65535 pulses (±1 pulse)
- Pulse width: 4 ms – 2000 s
Specifications for Mammo
- Time: 0.1 ms – 2000 s (±0.5 % or ±0.5 ms)
- Dose: 1 nGy – 1000 Gy. 3 µR – 150 kR (±5 %)
- Dose Rate: 25 nGy/s – 530 mGy/s (±5 % or ±12 nGy/s)
- HVL Range: 0.19 – 4.3 mm Al (depending on Target/Filter combination) (±10 %)
- Pulses: 1 – 65535 pulses (±1 pulse)
Specifications for CT
- Tube Voltage: 45 – 155 kV (±1.5 %)
- Time: 0.1ms – 2000 s (±0.5 % or ±0.5 ms)
- HVL: 0,72 – 13 mm Al (±10 % or ±0.2 mm)
- Total Filtration: 1.0 – 90 mm Al (±10 % or ±0.3 mm)
- Pulses: 1 – 65535 pulses (±1 pulse)
Specifications for Dental
- Tube Voltage: 35 – 125 kV (±1.5 %)
- PPV: yes
- Time: 0.1ms – 2000 s (±0.5 % or ±0.5 ms)
- Dose: 1.3 nGy – 650 Gy. 150 nR – 74 kR. (±5 %)
- Dose Rate: 15 nGy/s – 320 mGy/s (±5 % or ±7 nGy/s)
- HVL: 0.72 – 13 mm Al (±10 % or ±0.2 mm)
- Total Filtration: 1.0 – 90 mm Al (±10 % or ±0.3 mm)
- Pulses: 1 – 65535 pulses (±1 pulse)
- Pulse rate/frequency: 0.5 – 180 Hz
- Pulse width: 4 ms – 2000 s
No reviews for this product